Products >> Portable flue gas analyzer Sensonic 1200
Sensonic 1200 Small, low cost hand-held combustion analyzer using electrochemical sensors. A full-blood analyzer despite compact size. This flue gas analyzer can be fitted with up to 3 electrochemical sensors. This is an attractive alternative to other, larger flue gas analyzers on the market. Is especially suited to short-term measurements on small burner systems. Designed according to EN50379. Pictures... Folder () Sensonic 1200
Standard and optional configuration
Measurement of gas concentrations
Standard version with two gas sensors: O2, CO
Optional gas sensor NO
CO ambient measurement with a resolution of 0.1 ppm
Measurement of other parameters
Measurement of gas and ambient temperatures
Pressure, draft and differential pressure measurements with 0.1 Pa resolution
Flow velocity measurement
Relative humidity measurement probe - optional
Calculation
CO2 concentration
Calculation of all relevant combustion parameters
Processing and presentation of measuring data
All results shown on display
Averaging of all measured values. Averaging time: 2, 10, 20, 30, 60, 120, 180 seconds
Graphical presentation of all measured values as diagram
Storage for 16 measurements - "measure now, print later"
Powerful PC program for flue gas analyzer settings and data communication
Software capabilities
Automatic zeroing when the analyzer is switched on
All settings for the analyser carried out with PC programm
List of 6 common fuels (from DIN or OENORM)
4 further freely programmable fuels
Permanent automatic check of the instruemnt with acoustic warning and full information in the "control list"
Hardware capabilities
Integrated clock/calendar
IR interface for external printer
Incredible 75 hours operating time from one set of batteries
Full graphic display LCD 66 x 36 mm
Gas probe with thermocouple and condensate trap
RS-232C interface and multifunctional PC program
Updateable FLASH ROM program memory
Simple, intuitive operation, like a mobile phone
Optional accessories
leather-look case for storage and rubber boot for protection of the instrument
Magnetic plate to attach the instrument to steel surfaces
External thermal printer with wireless IR communication
Technical data
Size
245 x 116 x 45 mm
Weight
0.6 kg
Probe length
150 mm
Length of gas line
2 m
In line filter
in-line, Filter 20 µm
Display
LCD Graphic display 36 x 66 mm
Power supply
6 V ( 4 x AA) Battery, or external power supply 9 VDC
Interface
RS232C for external computer
IR connection for external printer
Gas pump
Membrane pump
Clock/calendar
Integrated clock/calendar
Operating temperature
10 °C ÷ 50°C
Storage temperature
-20 °C ÷ +55 °C
Measurement
Parameter
Measuring method
Range
Resolution
Accuracy
Detection
Limit
Time (T90)
O2 - oxygen, volumetric concentration
electrochemical gas sensor
0..25 %
0.01 %
±0.2 % or 2 % rel.
0.01%
45 s
CO2 - carbon dioxide, volumetric concentration
calculated from volumetric concentration of O2
0..25 %
0.01 %
±0.2 % or 2 % rel.
0.01%
45 s
CO - ambient carbon monoxide, volumetric concentration
electrochemical gas sensor
0..100 ppm
0.1 ppm
±0.5 ppm or
5 % rel.
0.1 ppm
45 s
CO - carbon monoxide, volumetric concentration
electrochemical gas sensor
0..20000 ppm
0.1 ppm or
1 ppm as set
±5 ppm or
5 % rel.
1 ppm
45 s
COmg - carbon monoxide, mass concentration
calculated from volumetric concentration of CO
0...
1 mg/Nm3
±10 mg/Nm3
or 5 % rel.
1 mg/Nm3
45 s
COrel - carbon monoxide, mass concentration relative to O2
calculated from volumetric concentration of CO and O2
0...
1 mg/Nm3
±10 mg/Nm3
or 5 % rel.
1 mg/Nm3
45 s
NO / NOx - volumetric concentration of nitrogen oxides.
electrochemical gas sensor
0..1000 ppm
0.1 or 1 ppm
set by user
±5 ppm or
5 % rel.
1 ppm
45 s
NOmg/NOxmg - mass concentration of nitrogen oxides
calculated from volumetric concentration of NO
0...
3 mg/Nm3
±10 mg/Nm3
or 5 % rel.
3 mg/Nm3
45 s
NOrel / NOxrel - mass concentration of nitrogen oxides relative to O2
calculated from volumetric concentration of NO and O2