| Standard and optional configuration |
|
Measurement of gas concentrations
|
- Gas concentrations measured with NDIR sensors: CO2, CH4
- Gas concentrations measured with electrochemical sensors: O2
- Toxic gas concentrations measured with electrochemical sensors: CO, NO, NO2, SO2, H2S, H2., Please ask for quote for other gases.
|
|
Measurement of other parameters
|
- Measurement of gas and ambient temperatures
- Pressure, draft and differential pressure measurements with 1 Pa resolution
- Soot test according to Bacharach with a pump flow of 1.63 l/min
- 6 analog inputs (2 current/voltage, 2 thermocouple, 2 thermistor)
|
|
Calculation
|
- CO2 concentration
- Calculation of absolute and relative mass concentrations & relative emissions
- Calculation of all relevant combustion parameters
|
|
Processing and presentation of measuring data
|
- All results shown on display
- Averaging of all measured values. Averaging time: 10 sec - 60 min
- Graphical presentation of all measured values as diagram
- Memory capacity for 1024 sets of data, organised into 10 data banks
- Up to 9 separate items in a data set
- User definition of the 9 items
- Memory capacity for 30 reports
- Data logger function for the analog inputs
- All measured values, stored values or displays can be printed on the internal printer.
- Powerful PC program for flue gas analyzer settings and data communication
|
|
Software capabilities
|
- International compatibility (language, date format etc.)
- Password protection for settings
- Automatic zeroing when the analyser is switched on
- Calibration of O2/CO2 during use
- All parameters programmable
- List of 22 common fuels
- 10 further freely programmable fuels
- Permanent automatic check of the instrument with acoustic warning and full information in the "control list"
- Compensation of cross sensitivity and temperature drift of gas sensors
|
|
Hardware capabilities
|
- Peltier Gas dryer
- Electronic regulation of the pump flow rate
- Integrated clock/calendar
- Internal 57 mm dot-matrix printer
- Power from rechargeable battery or mains
- Big (75 x 64 mm) LCD display with backlighting (Graphics or 11 x 21 characters)
- Gas probe with thermocouple
- RS-232C interface and multifunctional PC program
|
|
Optional accessories
|
- External ambient temperature sensor
- Flow velocity
|
|
Measurements
|
| Parameter |
Method |
range |
Display resolution |
Accuracy |
Detection limit |
Response time (t90) |
| Gases measured in standard configuration |
| O2 - oxygen, volumetric concentration |
electrochemical gas sensor |
0…25 % |
0.01% |
±0.2 % or 2 % rel. |
0.2 % |
45 s |
| CO2 - carbon dioxide, volumetric concentration |
calculated from volumetric concentration of O2 |
0..25 % |
0.01% |
±0.2 % or 2 % rel. |
0.2 % |
45 s |
| CO - carbon monoxide, volumetric concentration |
electrochemical gas sensor |
0….20,000 ppm |
1ppm |
±5 ppm or 5 % rel. |
5 ppm |
45 s |
| COmg - carbon monoxide, mass concentration |
calculated from volumetric concentration of CO |
0... |
1 mg/Nm3 |
±10 mg/Nm3 or 5 % rel. |
10 mg/Nm3 |
45 s |
| COrel - carbon monoxide, mass concentration relative to O2 |
calculated from volumetric concentration of CO and O2 |
0... |
1 mg/Nm3 |
±10 mg/Nm3 or 5 % rel. |
10 mg/Nm3 |
45 s |
| Gases measured with optional IR sensors |
| CO2 - carbon dioxides volumetric concentration |
IR sensor
|
0...25 % 0...100 % |
0.01 % 0.1 % |
0.5 % from range or +/-3 % rel. |
0.2 % |
45 s |
| CH4 - Methane, volumetric concentration |
IR sensor |
0...5 % 0...100 % |
0.01 % 0.1 % |
0.5 % from Range or +/-3 % rel. |
0.2 % |
45 s |
| Gases measured with optional electrochemical sensors |
| NO / NOx - volumetric concentration of nitrogen oxides. |
electrochemical gas sensor
|
0...5000 ppm |
1 ppm |
± 5 ppm or 5 % rel. |
1 ppm |
45 s |
| NOmg/NOxmg - mass concentration of nitrogen oxides |
calculated from volumetric concentration of NO |
0... |
1 mg/Nm3 |
±10 mg/Nm3 or 5 % rel. |
1 mg/Nm3 |
45 s |
| NOrel / NOxrel - mass concentration of nitrogen oxides relative to O2 |
calculated from volumetric concentration of NO and O2 |
0... |
1 mg/Nm3 |
±10 mg/Nm3 or 5 % rel. |
1 mg/Nm3 |
45 s |
| NO2 - volumetric concentration of nitrogen dioxide. |
electrochemical gas sensor |
0...1000 ppm |
1 ppm |
±5 ppm or 5 % rel. |
1 ppm |
45 s |
| NO2mg - mass concentration of nitrogen dioxide |
calculated from volumetric concentration of NO2 |
0... |
1mg/Nm3 |
±15 mg/Nm3 or 5 % rel. |
2 mg/Nm3 |
45 s |
| NO2rel - mass concentration of nitrogen dioxide relative to O2 |
calculated from volumetric concentration of NO2 and O2 |
0... |
1mg/Nm3 |
±15 mg/Nm3 or 5 % rel. |
2 mg/Nm3 |
45 s |
| SO2 - volumetric concentration of sulfur dioxide. |
electrochemical gas sensor |
0...5000 ppm |
1 ppm |
±5 ppm or 5 % rel. |
1ppm |
45 s |
| SO2mg - mass concentration of sulfur dioxide. |
calculated from volumetric concentration of SO2 |
0... |
1 mg/Nm3 |
±5 mg/Nm3 or 5 % rel. |
3 mg/Nm3 |
45 s |
| SO2rel - mass concentration of sulfur dioxide. relative to O2 |
calculated from volumetric concentration of SO2 and O2 |
0... |
1 mg/Nm3 |
±15 mg/Nm3 or 5 % rel. |
3 mg/Nm3 |
45 s |
| Other measured values |
| Tgas - flue gas temperature |
Thermocouple
|
-10...1000°C |
1°C |
± 2 °C or 1. 5 % rel. |
1 °C |
30 s |
| Tamb - ambient temperature |
Thermistor |
-10...100°C |
1°C |
± 1 °C |
1 °C |
30 s |
| U1,U2 - 2 external inputs |
ADC
|
-20...+20 V
|
0.01 V
|
±0.02 V
|
0.01 V
|
10 s
|
| I1,I2 - 2 external inputs |
ADC
|
-20...+20 mA
|
0.01 mA
|
±0.02 mA
|
0.01 mA
|
10 s
|
| T1,T3 - 2 external inputs |
Thermocouple
|
0...1600°C
|
1°C
|
±2 °C or 1.5 % rel.
|
1 °C
|
10 s
|
| T2,T4 - 2 external inputs |
Thermistor
|
-20...100°C
|
1°C
|
±2 °C or 1.5 % rel.
|
1 °C
|
10 s
|
|
| Pressure/diff. pressure |
DMS bridge |
-25 hPa ... +25 hPa |
1 Pa |
±2 Pa or 5 % rel. |
1 Pa |
10 s |
| flow velocity (Option) |
Pitot tube |
1...50 m/s |
0.1 m/s |
0.3 m/s or 5 % rel. |
0.1 m/s |
10 s |
| Soot/smoke test (option) |
Bacharach method |
0...9 |
0.5 |
0.5 |
0.5 |
|
|
| TI (CO/CO2-Toxic Index) |
calculated |
0...0.01 |
0.0001 |
5 % rel. |
0 |
10 s |
| Lambda - excess air number |
calculated |
1...10 |
0.01 |
2 % rel. |
0 |
10 s |
| qA - combustion losses |
calculated |
0...100% |
0.1% |
2 % rel. |
0 % |
10 s |
| Eta - efficiency |
calculated
|
0..120 %
|
0.1 %
|
2 % rel.
|
0 %
|
10 s
|